| Model |
AP-300 |
Cat.No. |
5291 |
| Measurement Scales |
Angle of Rotation,
International Sugar Scale (without temperature compensation),
International Sugar Scale (with Automatic Temperature Compensation) |
| Measurement Readings |
Angle of Rotation,
International Sugar Scale (without temperature compensation),
International Sugar Scale (with Automatic Temperature Compensation ),
Specific Optical Rotation, Concentration, and Purity |
| Measurement Range |
Angle of Rotation
-89.99 to +89.99°
International Sugar Scale
-130.00 to +130.00°Z |
Minimum Indication |
Angle of Rotation 0.01°
International Sugar Scale
0.01°Z |
| Measurement Accuracy |
Angle of Rotation
Displayed value ±0.01° (- 35.00 to + 35.00° )
Relative precision ±0.2% (- 35.01° to - 89.99° , + 35.01° to + 89.99° )
International Sugar Scale
Displayed value ±0.03°Z (- 101.00 to + 101.00° )
Relative precision ±0.2% (- 130.00 to - 101.01°Z , + 101.01 to + 130.00°Z )
( Checked by reading a standard Quartz Plate ) |
| Temperature correction range for International Sugar Scale |
18.0 to 30.0° C |
Manual Calibration |
Enables manual calibration by measuring a standard Quartz Plate |
| Measurement Wavelength |
589nm(D-line) |
Display Panel |
Color LCD with backlight |
| Power Supply |
AC100V to 240V 50/60Hz |
| Power Consumption |
50VA |
Output Terminals |
•Digital Printer DP-63 or DP-AD ( Optional )
• RS-232C |
| Dimensions & Weight |
48.5×28.5×17.5cm, 13.9kg
(Main Unit only) |
Accessories |
100mm (5ml), 200mm (10ml)
Observation tube (1each) |